top of page
TK Electrolyte Dispenser

The Ormity CTI Drop Dispenser is a precision dispensing module designed for controlled drop generation in Comparative Tracking Index (CTI) tests in accordance with IEC 60112.

The system utilizes a single-channel electronic pipette mounted on a rigid mechanical holder, actuated by a micro-stepping motor (NEMA 12) through a PTFE actuator tip. This ensures repeatable, hands-free operation and consistent drop release, eliminating operator variability.

 

Note: The dispenser is intended for integration within CTI test equipment - Dunel or Ormity. Volumetric performance is defined by the electronic pipette; the module serves as the mechanical actuation interface.

 

Applications & Compliance

 

  • Comparative Tracking Index (CTI) testing.

  • Proof Tracking Index (PTI) testing.

  • Electrical insulation material evaluation per IEC 60112.

  • Compatible with ISO/IEC 17025 laboratory practices.

  • Volume verification traceable to ILAC MRA (via accredited pipette calibration).

 

Technical Specifications

Dispensing Characteristics

ParameterSpecification
Nominal Drop Volume20 µL
Verification Volume25 µL
Drops per Test50
Total Test Volume≈ 1.0 mL
Drop IntervalAs specified by IEC 60112 (e.g., 30s ± 5s)
Dispensing ModeSingle drop per actuation

 

 

Actuation System

ParameterSpecification
Actuator TypeMicro-stepping motor
Motor SizeNEMA 12
Actuation MethodLinear push on pipette button
Actuator TipPTFE (Teflon™) for low friction
Control InterfaceExternal controller / Automated test system

 

Pipette Requirements

To ensure system integrity, the pipette used must meet the following criteria:

  • Type: Single-channel electronic pipette.

  • Volume Range: 5 – 50 µL.

  • Resolution: 0.1 µL or better.

  • Calibration: Independently calibrated and verified.

  • The pipette is not supplied as part of the actuator module unless explicitly specified.

 

Mechanical & Environmental Features

  • Construction: Rigid aluminum column for precise vertical alignment.

  • Mounting: Custom mechanical holder with adjustable stroke alignment.

  • Material Integrity: Chemical-resistant materials in contact areas; designed to prevent lateral loads on the pipette piston.

  • Operating Temperature: 20 ± 5 °C.

  • Relative Humidity: ≤ 75 % RH (non-condensing).

 

Volumetric Performance

Performance values are dependent on the specific electronic pipette utilized.

Parameter (at 25 µL)Typical Value
Systematic Error± 0.25 µL (± 1.0 %)
Random Error± 0.25 µL (± 1.0 %)

 

 

Maintenance & Calibration

  • Pipette: Shall be calibrated separately according to the laboratory’s quality system (recommended every 12 months).

  • Mechanical Module: The holder and actuator are non-volumetric components and do not require liquid-handling calibration.

  • Cleaning: Use isopropyl alcohol or mild laboratory detergent on external surfaces.

 

Manufacturer: Ormity Inc.

TK Electrolyte Dispenser

SKU: DNTK112-50LSI
€3,200.00Price
Quantity
    bottom of page