CTI Electrolyte Dispenser – IEC 60112

CTI drop dispenser with motor-actuated electronic pipette for tracking tests, per IEC 60112.

 

Standards

  • IEC 60112 – CTI / PTI drop application

 

Technical data

Nominal drop volume20 µL (verification 25 µL)
Drops per test50 (≈ 1.0 mL)
Drop interval30 s ± 5 s
ActuatorMicro-stepping motor NEMA 12, PTFE tip
Pipette requiredSingle-channel electronic, 5–50 µL, resolution ≤ 0.1 µL, calibrated
Systematic / random error at 25 µL± 0.25 µL (± 1 %) each, pipette-dependent
Environment20 ± 5 °C, ≤ 75 % RH
ColumnRigid aluminum

 

Ref.: DNTK112-50LSI (Dunel) / ORTYTK112-50LSI (Ormity) – former codes: OTY112-500-SLI-Y